N103A PRECISION ANALOG TENS
2- Facilidades de pago
- 14-días para retornar
- Entrega: 1-2 Días
- Indicated for the treatment of pain (acute or chronic) such as neck pain, low back pain, epicondylitis, rheumatism, sciatica
- The device is provided with two output channels controllable, each one in a way independent . Two electrodes can be connected to each output channel
- The N103A electronics create electrical impulses whose intensity, duration, number per second and modulation can be modified with the control buttons
- The controls are very easy to use and the sliding lid prevents accidental changes to settings
Content from the box:
- 1 unit
- 4 electrodes 4x4 cm
- 2 cables (KE24)
- 1 9V battery
- 1 manual
- 1 briefcase
He Precision TENS N103A carry a microprocessor built-in, which improves the precision of TENS
|
MECHANISM |
TECHNICAL DESCRIPTION |
01 |
Channel |
Dual, independent between channels |
02 |
pulse width |
Adjustable, 0-90 mA, peak at 500 ohm |
03 |
Voltage |
Adjustable, 0-100 volts |
04 |
Pulse rate |
Adjustable, from 2 to 150Hz |
05 |
Pulse width |
Adjustable, from 50 to 300 microseconds |
06 |
Mode |
4 modes, B (Burst), C (constant), M (Width Modulation), MM (Evaluate and Width Modulation) |
07 |
Size |
95(L)x65(W)x23.5(H)mm |
08 |
Weight |
115 grams (battery included) |
09 |
Burst Mode (B) |
Pulse width (adjustable), frequency = 100 Hz |
10 |
Constant mode (C) |
Pulse rate and pulse width are adjustable. Generates continuous stimulation based on an adjustable value |
eleven |
Modulation Mode (M) Width Modulation |
The pulse width decreases from the preset value by 60% in 3 seconds and then returns to the original setting. Total cycle time is 6 seconds. Pulse frequency (2 ~ 150Hz) and pulse width (50 ~ 300μs) are fully adjustable |
12 |
Modulation Mode (MM) Evaluate and Width Modulation |
Modulation Mode is a combination of pulse frequency and pulse width modulation. Pulse rate and width are automatically varied in a cycling pattern. The pulse width is reduced by 50% of its original setting in 0.5 seconds, then the pulse rate is reduced by 50% of its original setting in 0.5 seconds. Total cycle time is 1 second. In this mode, the pulse
|
13 |
waveform |
Asymmetric biphasic square pulse |
14 |
Timer |
30, 60 minutes or Continue (C) |
fifteen |
Max. Pulse charge |
20 microcoulombs |
16 |
Power supply |
A 9 volt battery. (Alkaline or rechargeable nickel-cadmium |
17 |
Battery duration |
Approximately 50 hours at nominal values. |
18 |
Funcionament condition |
Temperature: 0 ℃ ~ 40 ℃
|
19 |
Observation |
There is maybe +/-10% tolerance of all parameters and/+-20% tolerance of voltage amplitude |